Foreach TechnologyForeach Technology
Home
Products
Applications
About Us
Contact & Partnership
简体中文EnglishEspañolFrançais한국어Русский
Home
Products
Pump SeriesValve SeriesTubing SeriesFitting SeriesProbe SeriesSmart Control Series
Applications
IVD DiagnosticsLife SciencesLaboratory AutomationAnalytical InstrumentsEnvironmental MonitoringSynthetic Biology
Resources
DatasheetsFluid Resistance CalculatorFitting ReplacementInstallation GuidesMaterial CompatibilityTechnical ArticlesCompany News
About Us
About ForeachR&D and ManufacturingQuality System and QualificationsMilestonesForeach Culture
Contact & Partnership
Contact UsBecome a Distributor
  • Pump Series
  • Valve Series
  • Probe Series
  • Fitting Series
  • Tubing Series
  • Control Series
  • IVD
  • Life Science
  • Lab Automation
  • Analytical Instruments
  • Environmental Monitoring
  • Synthetic Biology
  • Datasheets
  • Fluid Resistance Calculator
  • Fitting Replacement
  • Installation Guides
  • Material Compatibility
  • Technical Articles
  • Company News
  • About Foreach
  • R&D & Manufacturing
  • Quality & Compliance
  • History
  • Foreach Culture
  • Inquiry Form
  • Contact Information

Shenzhen Foreach Technology Co., Ltd.

Address: 13th Floor, 2nd Building, Yufengda Industrial Park, 1008th Guang Qiao Boulevard, Guangming District, Shenzhen 518132 CHINAView Map

Email: sales@foreachtek.com

Tel: +86 755 8655 3831

LinkedInYouTubeFacebookInstagramX

© 2026 Shenzhen Foreach Technology Co., Ltd. All rights reserved.

Privacy Policy·

Home/Product Center/Miniature Diaphragm Pumps/DPL30H-DS
Foreach DPL30H-DS 600 kPa Brushed DC High-Pressure Miniature Liquid Diaphragm Pump, 3,000 h Service Life, 12 V / 24 V

Foreach DPL30H-DS 600 kPa Brushed DC High-Pressure Miniature Liquid Diaphragm Pump, 3,000 h Service Life, 12 V / 24 V

The Foreach DPL30H Series Brushed High-Pressure Miniature Liquid Diaphragm Pump is designed for liquid transfer under higher-pressure conditions, with a no-load flow rate of 300 mL/min, a pressure rating of 600 kPa, a brushed motor, and a service life of 3000 h. It is suitable for automated instruments with higher tubing resistance, higher back pressure, or stable liquid-transfer requirements, and can be selected for IVD, life science, laboratory automation, and analytical instrument applications. Actual performance may vary with the medium, tubing, and operating pressure. Refer to the specification table below for detailed parameters. The stated service-life value applies at rated voltage under continuous operation.

Typical Applications:

High-back-pressure liquid transfer, cleaning-solution transfer, reagent transfer, liquid circulation

Model:This product is custom-made
Motor TypeBrushed DC Motor
Product TypeHigh-Pressure Miniature Liquid Diaphragm Pump
Rated VoltageDC 24V ±10% / DC 12V ±10%
Rated Power≤7 W
Free-Flow Rate300 mL/min
Rated Pressure600 kPa
Self-Priming Height3 mH₂O
Working MediumPurified water; other fluids require evaluation
Fluid Temperature+5 °C–+40 °C
Port SizeThreaded Port, connects to 6 × 4 mm rigid tubing
Noise≤80 dB
Operating Temperature+5 °C–+40 °C
Operating Relative Humidity30%–85% RH (Non-Condensing)
Pump Head MaterialPP S
Diaphragm MaterialEPDM / PTFE
Valve MaterialEPDM / FFKM
Service Life3,000 h

Typical Applications for a 300 mL/min 600 kPa High-Pressure Miniature Liquid Diaphragm Pump

A 300 mL/min high-pressure miniature liquid diaphragm pump is intended for high-resistance fluid paths that conventional low-pressure diaphragm pumps may not be able to serve, including narrow-bore tubing, filters, multiple valves, flow cells, needles and other components that create substantial pressure drop. The DPL30H has a free-flow rate of 300 mL/min, a rated pressure of 600 kPa and a self-priming height of 3 mH₂O for instrument fluid transfer requiring higher delivery pressure.

The 300 mL/min free-flow rate and 600 kPa rated pressure are not the same operating point. Selection must be based on the target flow, inlet conditions, discharge back pressure and total fluid-path resistance; it must not be interpreted as maintaining 300 mL/min at 600 kPa.

Depending on its function in the equipment, the same pump may be described as a high-pressure wash-solution pump, high-back-pressure transfer pump, rinse pump, auxiliary supply pump or circulation pump for a high-resistance fluid path.

1. High-Resistance Wash-Solution Pump for IVD, Biochemical and Analytical Instruments

In IVD, biochemical and other analytical instruments, wash solution may pass through solenoid valves, manifolds, filter assemblies, probes or narrow internal tubing before reaching the cleaning point. When several local restrictions are connected in series, the required pressure can be substantially higher than in an open, low-resistance transfer path.

In this type of circuit, the DPL30H can serve as a wash-solution or rinse pump for a high-resistance cleaning path. Whether the 600 kPa platform is required should be determined from the actual pressure drop at the target wash flow, not from the application label IVD washing alone.

2. Long-Line and Multi-Valve Liquid Transfer in Laboratory Automation

Laboratory automation equipment may contain a reservoir, solenoid valves, distribution manifolds, sensors, filters and multiple fluidic modules. Pressure losses accumulate when tubing is long or narrow, or when several valves and fittings are connected in series.

The DPL30H can be evaluated at the required operating point for auxiliary supply or washing circuits that must overcome this higher system resistance. The 300 mL/min value is a free-flow reference; installed flow still depends on tubing diameter and length, valve pressure drop, filter condition and discharge pressure.

3. Supply Upstream of Filters and Flow Cells in Analytical Instruments

Some analytical instruments must transfer liquid through filters, flow cells or other internal components with narrow passages. As filter resistance increases, or as the bore of the flow cell and connecting tubing decreases, the pump must generate a larger differential pressure to maintain the target flow.

The DPL30H can support non-metering auxiliary supply, washing and fluid-replacement duties in these circuits. Filter pressure drop may change from a new element to the end of its service interval, so selection should include the worst-case operating point after filter loading rather than only the initial condition.

4. Pressurized Transfer Through Needles, Narrow Tubing and Small Outlets

Needles, capillaries, small-ID rigid tubing and other small outlets increase flow resistance, and the resulting pressure drop changes significantly with flow rate, fluid viscosity, bore and effective length.

When an instrument must perform washing, auxiliary supply or fluid replacement through these high-resistance components, the DPL30H can be evaluated against the required flow and calculated or measured pressure demand. Its high-pressure capability is used to overcome fluid-path resistance and does not make it a high-accuracy metering pump.

5. Liquid Circulation and Auxiliary Transfer with Higher Back Pressure

Some instrument circuits must deliver wash solution, buffer or another compatible liquid to a downstream module that is already under pressure, or circulate liquid periodically through a loop containing multiple valves, sensors and fluidic components.

In these systems, the operating point is governed by the pressure difference between the inlet and outlet together with total circuit resistance. If a conventional 100 kPa diaphragm pump is already close to or beyond its applicable range at the target flow, the DPL30H 600 kPa platform can be evaluated, but final selection must be based on a complete fluid-path test and a valid flow-pressure curve.

Brushed vs Brushless High-Pressure Miniature Liquid Diaphragm Pump: Which Should You Choose?

Brushed Motor

The brushed DPL30H version uses a brushed DC motor and has a specified service life of 3,000 h. It is suited to cost-sensitive instruments with a defined duty cycle where high-pressure transfer is not required for continuous long-term operation.

Mechanical brushes remain in contact with the commutator inside the motor, so extended operation produces brush-wear particles and mechanical commutation may generate arcing and transient electrical noise. If the complete instrument has stricter maintenance, internal-cleanliness or EMC requirements, the brushless version should be evaluated.

Brushless Motor

The brushless DPL30H version uses electronic commutation, with no mechanical brush wear or brush-commutation arcing. Its specified service life is 10,000 h, making it better suited to high-resistance fluid systems that require longer operating periods or extended maintenance intervals.

Brushless operation does not eliminate electromagnetic interference. Electronic commutation and the driver circuitry can still generate electromagnetic noise, so final EMC performance must be verified with the complete power supply, wiring, grounding, shielding and control system.

Compare with the DPL30H Brushless High-Pressure Liquid Diaphragm Pump →

Application Selection Note

For the DPL30H, 300 mL/min is the free-flow rate and 600 kPa is the rated pressure; they are not the same operating point. Define the target flow first, then calculate the total pressure differential created by narrow tubing, valves, filters, flow cells, needles, discharge pressure and inlet conditions, and confirm the actual operating point with a valid flow-pressure curve.

The DPL30H is intended for high-resistance or higher-back-pressure circuits that conventional low-pressure diaphragm pumps may not be able to serve. Final validation must also consider medium compatibility, duty cycle, inlet conditions, system back pressure, ambient temperature and equipment-life requirements.

Learn more: Operating point and selection of the DPL30H high-pressure miniature liquid diaphragm pump →
Preview DrawingView the technical drawing for Foreach DPL30H-DS 600 kPa Brushed DC High-Pressure Miniature Liquid Diaphragm Pump, 3,000 h Service Life, 12 V / 24 V.

Datasheet not yet available

The datasheet for this product has not been uploaded yet.

Frequently Asked Questions

DPL30H is a high-pressure miniature liquid diaphragm pump with a no-load flow rate of 300 mL/min and a pressure rating of 600 kPa. The standard DPL30 has the same no-load flow rating and a pressure rating of 100 kPa. DPL30H is more suitable for fluid paths with higher back pressure or tubing resistance.

It is suitable for cleaning-solution transfer, reagent transfer, and liquid circulation where higher tubing resistance, longer fluid paths, or higher back pressure must be overcome.

Yes. Actual flow may decrease as operating pressure and fluid-path resistance increase, so testing should be performed under the equipment’s actual pressure conditions.

The brushed version has a service life of 3000 h. Actual service life may also be affected by load, start-stop frequency, and operating environment.

Consider DPL30H when the standard DPL30 cannot overcome fluid-path resistance or when the equipment has a higher back-pressure requirement.

Related Videos

Related Technical Articles

Foreach miniature diaphragm pumps for OEM integration: operating points, instrument validation and production introduction

Whether the miniature diaphragm pump can work stably in OEM instruments depends on the joint matching of fluid path tasks, inlet and outlet conditions, material combinations, control methods and overall machine structure. Taking Foreach DPL30, DPL60, DPL30H and DPGL800 as examples, this article starts from the flow and pressure budget, gas-liquid status, wetted materials and electrical interfaces to explain how to establish candidate solutions, carry out reproducible prototype verification, and convert the verification results into technical requirements for production supply.

2026-09-11View Details

Selecting a Self-Priming Miniature Liquid Diaphragm Pump: Lift, Priming Time and Dry or Wet Starts

Select a self-priming miniature liquid diaphragm pump by checking suction lift, time to first liquid, dry and wet starting conditions, and stable delivery. Use DPL30, DPL60 and DPL30H specifications to define an instrument-level test.

2026-09-07View Details

Reducing Flow Pulsation in Miniature Diaphragm Pumps: Measurement and Practical Measures

Diagnose miniature diaphragm pump flow pulsation by separating real fluctuations from measurement errors. Compare damping, tubing and speed-control measures using matched operating conditions and endpoint measurements.

2026-09-07View Details

Miniature Diaphragm Pump Noise: Causes, Vibration Isolation and Noise Reduction

Investigate miniature diaphragm pump noise at a matched operating point. Check the motor, pump head, supports, tubing and enclosure, then validate noise-reduction measures against flow, temperature rise and reliability.

2026-09-07View Details

Miniature Diaphragm Pump Backflow After Stopping: Check Valves and Anti-Siphon Design

Diagnose reverse flow after a miniature diaphragm pump stops, distinguish it from forward siphoning and residual dripping, and evaluate check-valve behavior, valve sequencing and static sealing in the complete circuit.

2026-09-07View Details

Which Materials Are Available for Miniature Diaphragm Pumps? Selection Considerations

Compare pump-head, diaphragm and valve materials, then check fluid composition, concentration, temperature, exposure time and cleaning conditions before selecting a configuration.

2026-09-07View Details

What Are Cleaning, Wash and Rinse Pumps in IVD Systems, and Why Are Miniature Diaphragm Pumps Used?

Cleaning pump, wash pump and rinse pump usually name different duties within an IVD cleaning sequence, not three fixed pump mechanisms. Selection starts with wash volume, cycle rate, line resistance, wetted materials, and whether the duty is liquid supply or waste aspiration.

2026-09-03View Details

Why Does Actual Flow Change So Much When the Tubing Size Changes on the Same Miniature Diaphragm Pump?

Even when the pump is unchanged, installed flow can change because of tubing inner diameter, length, bends, and the minimum bore of fittings. When flow falls after a tubing change, the effective response is neither to blame the pump first nor to enlarge every tube indiscriminately. Measure inlet pressure, outlet pressure, and accumulated flow together to determine where the added restriction actually occurs.

2026-08-24View Details

Which Has a Greater Effect on a Miniature Diaphragm Pump: Suction-Line or Discharge-Line Resistance?

Pressure loss on either the suction or discharge side increases the differential pressure carried by a miniature diaphragm pump, but there is no universal rule that one side is always more important. The suction side more readily affects initial priming, chamber filling, sensitivity to air leakage, outgassing, and cavitation. The discharge side more often appears as increased backpressure, reduced flow, and deadheading. Diagnose the system by observing inlet absolute pressure, outlet pressure, and flow on the same time axis.

2026-08-21View Details

What Is the Difference Between Rated Working Pressure, Maximum Output Pressure, Proof Pressure, and Burst Pressure for a Miniature Diaphragm Pump?

A pressure value has engineering meaning only when it is paired with a term definition, test fluid, inlet condition, corresponding flow rate, duration, and acceptance criteria. Rated working pressure can define an operating boundary; maximum output pressure, proof-test pressure, and burst pressure cannot be treated directly as continuous operating points.

2026-08-18View Details

Is 100 kPa Enough? How to Determine Whether a Fluid Path Needs a 600 kPa High-Pressure Miniature Diaphragm Pump

100 kPa and 600 kPa are not simply a standard version and an upgraded version. First complete a pressure budget for the entire fluid path at the target flow, then read the candidate pump curve. A higher pressure rating has value only when the application truly has high backpressure.

2026-08-15View Details

How Do You Build a Pressure Budget for a High-Backpressure Fluid Path? A Foreach 600 kPa Miniature Diaphragm Pump Selection Example

For a high-backpressure fluid path, do not tell a supplier only that you need 6 bar. Break down steady-state pressure losses at the target flow, establish normal steady-state, worst steady-state, pressure-margin, and transient-protection boundaries separately, and then compare them with the candidate pump's controlled curve.

2026-08-12View Details

How to Select a High-Pressure Diaphragm Pump: DPL30H 300 mL/min and 600 kPa Guide

Filters, narrow rigid tubing, valves, mixing chambers and long lines all create pressure drop. As backpressure rises, a pump's delivered flow falls, so no-load flow alone is not a valid selection criterion. DPL30H is designed for liquid paths that need greater pressure margin: 300 mL/min no-load flow and 600 kPa rated pressure.

2026-07-19View Details

How to Read a Diaphragm Pump Flow-Pressure Curve—and Predict Installed Flow

A catalogue value such as 300 or 600 mL/min is not a fixed flow rate inside an instrument. The useful engineering question is: at the actual inlet pressure, outlet backpressure, fluid, tubing and supply voltage, where will the pump and the fluidic circuit settle?

2026-07-28View Details

2-Wire vs 5-Wire Brushless Diaphragm Pumps: What Is the Difference and Which Should You Choose?